便携式X光机操作方法
便携式X光机操作方法
Portable X optical machine operation method
1、使用前首先检查仪器箱外观在运输过程中有无损坏。开箱检查主机是否有碰撞损伤痕迹。然后取出主机,将遮光盘的三个缺口对准显示屏前端的三个螺钉推入,旋转一角度即可使遮光盘固定。
1, before use, first check the appearance of the instrument box in the transportation process whether there is any damage. Open the box to check whether there are any collision damage marks on the main engine. Then take out the main machine, put the three notches of the disc on the front of the display screen three screws into, rotate an angle to make the disc fixed.
2、接通整机电源(将电源插头插入具有接地线的三孔单相电源插座),开关电源指示灯亮,主机数显表头亮。按下自锁启动按钮,数显表显示出X射线管阳极电压(KV)或阳极电流(μA)(视选择转换开关而定),说明仪器已正常工作,此时荧光屏呈黄绿色,将被测物放在X射线管与像增强器之间,(物体尽量靠近像增强器),即可在荧光屏上看到被测物的透视图象,调节数显面板上的电流(μA),电压(KV)可调旋钮,直到图象的清晰度满意为止。在保证图象清晰的前提下,应尽可能采用低电流(μA)进行透视。
2. Turn on the power supply of the whole machine (insert the power plug into the three-hole single-phase power socket with the grounding wire), the switch power indicator light is on, and the main digital display meter head is on. Pressing the self-locking start button, the digital display displays either the anode voltage (KV) or the anode current (mu A) of the X-ray tube (depending on the selection of the switch), indicating that the instrument is working properly. At this time, the fluorescent screen is yellow-green, and the object is placed between the X-ray tube and the image intensifier (the object is as close as possible to the image intensifier). On the screen to see the measured object perspective image, adjust the digital display panel on the current (mu A), voltage (KV) adjustable knob until the image resolution is satisfactory. Under the premise that the image is clear, low current (A) should be used for fluoroscopy.
3、当手按下自锁启动按钮时,X射线管发出射线,在数显面板上有工作指示灯指示,再次按动自锁启动按钮或定时时间到,X射线管停止工作(定时设定在1分钟左右)。4、本仪器宜间断使用,若连续工作时间过长,超过本机有效工作温度范围,本机能自动关机,此时在数显板上超温指示灯亮,需停机片刻待温度降低后才能恢复工作,停机时间视环境温度而定。
3. When the self-locking start button is pressed by hand, the X-ray tube emits rays, and there is a working indicator on the digital display panel. When the auto-locking start button is pressed again or the timing time arrives, the X-ray tube stops working (the timing is set at about 1 minute). 4. The instrument should be used intermittently. If the continuous working time is too long and exceeds the effective working temperature range of the machine, the machine can shut down automatically. At this time, the over-temperature indicator light on the digital display board is on, and the shutdown time depends on the ambient temperature.