便携式X光机操作方法
便携式X光机操作方法
Portable X optical machine operation method
1、使用前首先检查仪器箱外观在运输过程中有无损坏。开箱检查主机是否有碰撞损伤痕迹。然后取出主机,将遮光盘的三个缺口对准显示屏前端的三个螺钉推入,旋转一角度即可使遮光盘固定。
1, before use, first check the appearance of the instrument box in the transportation process whether there is any damage. Open the box to check whether the host has a collision damage trace. Then the host is removed, and the three gaps of the optical disk are pushed into the three screws on the front end of the display screen, and the optical disk is fixed by rotating one angle.
2、接通整机电源(将电源插头插入具有接地线的三孔单相电源插座),开关电源指示灯亮,主机数显表头亮。按下自锁启动按钮,数显表显示出X射线管阳极电压(KV)或阳极电流(μA)(视选择转换开关而定),说明仪器已正常工作,此时荧光屏呈黄绿色,将被测物放在X射线管与像增强器之间,(物体尽量靠近像增强器),即可在荧光屏上看到被测物的透视图象,调节数显面板上的电流(μA),电压(KV)可调旋钮,直到图象的清晰度满意为止。在保证图象清晰的前提下,应尽可能采用低电流(μA)进行透视。
2, turn on the power of the whole machine (insert the power plug into three hole single phase power socket with grounding line), switch power indicator light is bright, and the host digital display head is bright. Press the self lock start button, the digital display table shows the anode voltage (KV) or anode current (A) of the X ray tube (depending on the selection switch), indicating that the instrument has been working normally, and the fluorescent screen is yellow green at this time, and the measured objects are placed between the X ray tube and the image intensifier (as close to the image intensifier), and the fluorescence can be used in the fluorescence. On the screen, you can see the perspective image of the object being measured, adjust the current (A) and the voltage (KV) adjustable knob on the digital display panel until the clarity of the image is satisfied. Under the premise that the image is clear, low current (A) should be used for fluoroscopy.
3、当手按下自锁启动按钮时,X射线管发出射线,在数显面板上有工作指示灯指示,再次按动自锁启动按钮或定时时间到,X射线管停止工作(定时设定在1分钟左右)。4、本仪器宜间断使用,若连续工作时间过长,超过本机有效工作温度范围,本机能自动关机,此时在数显板上超温指示灯亮,需停机片刻待温度降低后才能恢复工作,停机时间视环境温度而定。
3, when the hands press the self locking button, the X ray tube sends out rays, and there is a work indicator on the digital display panel. Again, the X - ray tube stops working (set at about 1 minutes). 4, this instrument should be used intermittently. If the working time is too long and exceeds the effective working temperature range of the machine, this function automatically turns off the machine. At this time, the supertemperature indicator light on the digital display board is bright. It needs to stop for a moment to restore the work after the temperature is reduced, and the shutdown time depends on the temperature of the environment.